User`s guide
78 | Dome CXtra
2 To use the settings in the DIN constancy specification
(recommended), click Restore Defaults.
NOTE: Using the default settings is recommended to assure
conformance with the DIN specification.
The default values for Veil Test Interval and the Contrast Ratio Test
Interval depend on whether Luminance Control and Contrast
Stability are shown as Present or Not Present. They also depend on
whether you enable Constant Ambient Illumination. (Luminance
Control always appears as Present for Dome CX and Dome EX
displays and Not Present for Dome QX displays. Contrast Stability is
always shown as Not Present for both display types.)
3 To use thresholds and test intervals other than those in the DIN
specification (not recommended), change them as desired in the
Measurement Tests and Visual Tests sections of the tab.
NOTE: Do not use thresholds greater than 50%.
4 To create a reminder entry in the history log and change the Dome
CXtra icon on the taskbar to warning status when a test interval is
exceeded, click the Remind Me box next to the test. (The warning
icon is shown on page 10.)
5 Click Apply to accept the values on the tab, or click OK to accept
the values and close the DIN Test Setup dialog.
If you used any non-default setting or value, a warning message
appears when you click Apply or OK.