White Level Calibration
Correction of the display white level can now be done via manual
conformance testing or with the White Level Calibration service.
A one-time, end-user calibration of the white level may be necessary
when the target luminance is outside the tolerance established for
proper display conformance, such as when the white level of a display
has failed a manual conformance test, or when two or more displays
in a multi-unit configuration show disparate white levels. An external
photometer is required.
NOTE: Privileged users and administrators can calibrate the
white level if a display fails the White Level test during the manual
conformance testing. Only administrators can calibrate the white
level via the White Level Calibration service.
Support for 10-Bit Displays
You can now run the Dome CXtra software on Dome 10-bit grayscale
displays. You can check for 10-bit capability (both current and
maximum supported) on the Properties tab of the Reporting dialog.
You can also view 10-bit patterns (horizontal ramp and vertical
ramp test patterns) using the Test Patterns service. And in manual
conformance testing, you can take all 1024 measurements for more
precise testing results.
Using Dome CXtra with Dome Dashboard
For best practice of using Dome Cxtra 6.x with Dome Dashboard 2.x,
observe the following requirement and recommendation.
When using Dome Dashboard 2.x, make sure that Dome CXtra 6.x
or later is installed on each CXtra workstation.
On the CXtra workstation, limit the retention of history to 365 days
(one year).
Single Dome CXtra Package
The separate packaging of the Enterprise Management Service has
been discarded. All Dome CXtra services, including the Enterprise
Management Service, are now available in a single CD package.