User`s guide

46 | Dome CXtra
Performing a Conformance Test
You can use DICal service to verify that your display conforms to
the DICOM standard.
NOTE: This procedure requires installation of the optional LCD or
LCD II photometer. (See page 8.)
To perform a conformance test
1 If the Privilege service has been installed, log on as a privileged
user. (See page 96.)
2 Open the Dome CXtra Services menu and select
DICal Conformance > Conformance Wizard.
3 Follow the instructions for using the photometer in each
Conformance Wizard window and click Next.
NOTE: Press Ctrl+C to abort the test at any time.
4 If the last DICal Conformance Wizard window indicates that the
display conforms to DICOM standards, click Next or Finish. Go to
step 5.
If the last DICal Conformance Wizard window contains a warning,
click Next or Finish to close the window. Perform a custom DICOM
calibration as described on page 44. After calibration, perform
another conformance test. Start at step 1.
5 If necessary, review the conformance test data and test events
according to instructions in the following section.
Reviewing Conformance Test Data
Whenever you perform a conformance test, the DICal service stores
the generated data in both graphical and tabular formats.
To view the most recent conformance data, open the Dome CXtra
Services menu and select DICal Conformance > Conformance History.
The Conformance History dialog opens with Graphs and Table tabs.
The Graphs tab appears initially by default.