User`s guide
44 | Dome CXtra
Performing a Custom DICOM Calibration
A custom DICOM calibration uses the actual luminance values of
your display to adjust the response curve. After you complete the
calibration, the new values are applied in place of the default
DICOM calibration whenever DICOM Calibration is selected on
the services menu.
NOTE: This procedure requires installation of the optional LCD or
LCD II photometer as described on page 8.
To perform a custom calibration
1 If you have one of the display models listed below, follow
instructions provided with the display to verify that the
contrast is set to the value shown.
2 If the Privilege service has been installed, log on as a privileged
user. (See page 96.)
3 Open the Dome CXtra Services menu. Select DICal Calibration >
Calibration Wizard.
4 Follow the instructions for using the photometer in each
Calibration Wizard window and click Next.
NOTE: Abort the calibration at any time by pressing Ctrl+C.
5 If necessary, verify that your display calibration conforms to
the DICOM standard. (See page 46 for instructions.)
NOTE: If you change panel brightness using the brightness control
on the display, you must repeat the preceding calibration steps.
Display Model Contrast
GX2MP 70%
GX2MP (JUN 2007 or later) 70%
PX191 50%
PX1910M (2007 or later) 60%
PX1910M 50%
PX212M 80%
PX212M (RoHS) 50%