User`s guide

RightLight Service | 25
After you finish with the Conformance Wizard, RightLight enters
the event in both the Reporting History log and the Conformance
History log. All RightLight history events from manual conformance
tests are recorded in the Reporting History log. RightLight provides
additional information on conformance events in graphical and
tabular formats in the Conformance History log.
If the manual conformance test produces a warning condition
because the white level is beyond the tolerance set on the
Manual Alerts tab (page 22), then you can correct the warning by
calibrating the white level via the conformance test. Select the
check box to correct the warning, and click Next. The wizard
executes automatically.
Alternatively, you can adjust the tolerance for the failing test to
bring the display into conformance. (See also “White Level
Calibration Service” on page 35.)
NOTE: White Level calibration during conformance testing is
available only for Dome E2, Dome E3, Dome E3c, Dome E3cHB, and
Dome E5 displays. You must be an administrator or a privileged
user to perform a manual conformance test and to calibrate the
white level during the manual conformance test.