User`s guide
24 | Dome CXtra
Before starting a manual conformance test, select Dome CX/EX
DICOM Calibration on the services menu.
NOTE: If you run a conformance test with another response function
(such as Dome CX/EX Exponential Calibration or Dome CX/EX Native
Calibration), the system presents a warning message. Continuing to
run the test causes the system to show the Dome CXtra icon in the
warning state.
To run a manual conformance test
1 Open the Dome CXtra Services menu. Select RightLight
Conformance > Manual Conformance Wizard.
2 Follow the Conformance Wizard instructions.