DICal Service
You can calibrate GX24, GX2MP Plus, GX2MP, and select desktop displays to conform with
the DICOM curve or with another calibration response function. With the optional LCD
photometer, you can also perform custom calibrations and conformance tests.
Configuring the service for display calibration
First, check the brightness and contrast of your display. You must use the brightness and
contrast settings for your particular display. Look up the values in the Dome CXtra User’s
Guide and check your display documentation for how to change the settings, if necessary.)
Then, select DICal Panel Configuration on the CXtra Services menu. If your panel comes up
preselected, then the DICal service is already configured. Otherwise, click Edit Type on
the dialog and choose your panel.
After configuration, only privileged users can run the Calibration Wizard for custom DICOM
calibration. (Default DICOM calibration is active when no custom calibrations are run.) Use
the Reporting service to review calibration events. Switch between Original calibration (for
running Windows applications) and DICOM calibration (for viewing grayscale images) via
the CXtra Services menu.
Use the DICal Panel Configuration dialog
to verify the model of desktop display,
edit the model type, select a different
calibration response function, or
characterize an unknown display.