Specifications

86 Dolby
®
DP600 Program Optimizer Manual
System Administrator Functions
Recalibrating the Touch Screen
YourDP600touchscreenwascalibratedduringthemanufacturingprocess,butyoumay
needtoperformarecalibrationifyouexperienceanyirregularities.
Torecalibratethetouchscreen,usethetouchscreeninterfaceasfollows:
1. Press
config inthesystem screen.
The
configscreenappears,asshowninFigure C 8.Inthisscreen,youcanrecalibrate
thetouchscreen,ifrequired.
2. Press
touchscreen intheconfig screen.
The
calibratebuttonappears,asshowninFig ure C 10.
FigureC10
Figure C-10 Calibrate Button
3. Pressthecalibratebutton.
Thecalibrationsetupappearswithasmalltargetat onecornerofthescreen,asshown
inFigure C11.Tocalibratethetouchscreen,proceedasfollowswithinapproximately
thirtysecondsfromthetimeitappears.Ifyoudonotproceedwithinthistimeperiod,
thecalibrationscreendisappears.
•Press
thefirsttargetforacoupleofsecondsuntilitchangestoacross.
•Pressthe nexttargetuntilitchangestoacross.
Continuetheaboveprocedurefortheadditionaltargets.
Whenthecalibrationiscomplete,the
calibratebuttonreappears,andyoucannow
moveontoanyotherscreen,asrequired.
FigureC11
Figure C-11 Calibration Setup Screen
Press target