Focus microscope on the attached calibration ruler and thenadjust the
mangnifi cation till the picture is clear. Take a picture for the selected
area when the dial is parallel to the crossing as below .
Double-click on the picture to enter into mea-
surement interface as below.
Click Calibration icon , and then move mouse
to the picture. Click on the beginning and
ending points across a known value. Now a
dialog will popup as below and you need to
enter the exact value of the measurement
sample into ‘Actual dimension’.
Then the software will automatically calibrate
the magnifi cation rate. Click ‘OK’ and the adju-
sted magnifi cation rate will show in the upper
right corner.