Culture Myograph System Model 202CM
5.1.3 Changing O-rings
Each locknut on the culture myograph chamber is equipped with rubber O-rings
(ø1.07 x 1.27mm) to keep the chamber tight. These O-rings will from time to time
need to be replaced. The procedure is illustrated in Fig. 5.4.
Unscrew the locknut to reveal the O-ring and remove it carefully using small
forceps or a similar tool.
Insert the new O-ring into the chamber hole and push it back until it makes
contact with the bottom of the hole.
Carefully insert the glass cannula through the locknut and place it in the
chamber hole. Gently tighten the locknut.
Note:
Be careful not to damage the glass cannula tips when disassembling the culture
myograph chamber.
Figure 5.4 Changing O-rings.