2.3.3 Setup and Control of Superfusion Flow
The superfusion circuit consists of an inow and an outow from the culture
myograph chamber. Both ows are driven by the same peristaltic pump, which leads
the superfusion buffer from the reservoir to the culture myograph chamber and nally
to a waste bottle. The small steel pipe on the culture myograph chamber cover is
connected to a sterile micro lter (0.20µm). The small pipe works as a breathing valve
to prevent over pressure in the culture myograph chamber.
The superfusion buffer reservoir is continuously aerated with a mixture of 95%
O
2
and 5% CO
2
. The gas passes through a sterile micro lter (0.20µm) to prevent
contamination of the sterile superfusion buffer.
Peristaltic Pump
Waste Bottle
Culture Myograph Chamber
Oxygen
Supply
Superfusion
Buffer
Micro Filters
Figure 2.4 Illustration of perfusion ow control