Instruction manual
4 World Precision Instruments
INSTRUMENT DESCRIPTION
Instrument Description
Front Control Panel
The DLS100 control panel is shown Fig. 2.
Range Buttons
Range LEDs
Activity LED
Alarm LED
Power LED
Output Button
Output Connection
Constant Voltage Controls
Constant Current Controls
Fig. 2–DLS100 front panel
Range Buttons and Range LEDs–Select the operating range by pressing one of
the range buttons. They are marked according to the full-scale current or voltage
available. The adjacent Range LED indicator illuminates to indicate the selected
range. The full-scale value of the selected current or voltage output range is
delivered when the signal level from the DS8000 is 10V. The polarity of the output