Specifications
Construction of Absolute Reectance Attachment
100 % light
Measurement
light
Polarizer
position
Sample
Integrating
sphere
M2
M2'
M1
M3
Ref
25
Absolute Reectance Attachment
ASR-3105 Absolute Reectance Attachment, 5°
ASR-3112 Absolute Reectance Attachment, 12°
ASR-3130 Absolute Reectance Attachment, 30°
ASR-3145 Absolute Reectance Attachment, 45°
206-16817
206-16100
206-15001
206-15002
P/N
Model
These accessories are intended for use with the MPC-2600 Sample Compartment, and require the BIS-3100
Sample Base Plate-Integrating Sphere Set (P/N 206-17059). At larger angles of incidence (30°, 45°), a
polarizer is also required.
• Measurement wavelength range: 300 to 800 nm
• Accuracy: with respect to 90 % reectance samples
Incidence angle 5°: ± 1.5 %
Incidence angle 12°: ± 1.0 %
Incidence angle 30°, 45°: ± 2.5 %
• 100 % level sample setting: The sample measurement optical path can be switched using the single-touch V-N method.
• Approximate sample size: 25 to 200 mm dia., or 20 to 150 mm square, up to 30 mm thick
(Note)
The BIS-3100 Sample Base Plate-Integrating Sphere Set (P/N 206-17059) is required for mounting these absolute specular
reectance attachments.
Large Polarizer Set, Polarizer Type I, II, III
This accessory is for performing measurements when an absolute reectance attachment is used, with no
impact from polarization properties.
The Polarizer Adapter Set (P/N 206-15693) is simultaneously required for Polarizers Type I, II, and III.
LRLR-1 Large Polarizer Set
PLR-1 Polarizer Type I
PLR-2 Polarizer Type II
PLR-3 Polarizer Type III
206-15694
206-13236-01
206-13236-02
206-13163
20 mm
18 mm
17 mm
10 mm
250 to 2500 nm
400 to 800 nm
260 to 700 nm
260 to 2500 nm
P/N Effective Diameter Wavelength Range
Description
Powdered Sample Holder
(P/N 206-89065-41)
This powdered sample holder is for attachment to an integrating sphere. It can
be attached to all integrating spheres.
• Capacity of 0.16 mL, 3 included
Analog Output Interface
(P/N 206-25233-91)
Allows analog output for monitoring a liquid chromatograph, etc. and can be
connected to an integrator.
• Analog output full scale:
100 mV / 2 Abs or 100 mV / 100 %T