• Micro-volume samples can be measured. (Minimum sample volume: 50 L to 100 L)
• Support for commercial microplates and micro pipettes (with 8-series micro cell).
• Up to 16 samples can be measured at a time (with 16-series micro cell).
• Number of cells:
6 on the sample side (temperature-controlled)
1 on the reference side (temperature not controlled)
• Temperature control range: 16 °C to 60 °C
• Temperature display accuracy: ± 0.5 °C
• Temperature control precision: ± 0.1 °C
• Ambient temperature: 15 °C to 35 °C
• Number of cells: One each on the sample and reference sides (temperature-controlled)
• Temperature control range: 7 °C to 60 °C
• Temperature display accuracy: ± 0.5 °C
• Temperature control precision: ± 0.1 °C
This attachment permits measurement of up to six sample cells under
constant-temperature conditions. Combination of this attachment and the
Kinetics mode provides measurement of temperature-sensitive enzyme kinetics
of one to six samples.
CPS-100 Cell Positioner, Thermoelectrically
Temperature Controlled
(P/N 206-29500-**)
Uses Peltier effect for controlling the temperatures of the sample and reference
sample. No thermostatic bath or cooling water is required, so the operation is
quite simple and easy.
TCC-100 Thermoelectrically Temperature-
Controlled Cell Holder
(P/N 206-29510-**)
8/16-Series Micro Multi-Cell
Cell Holders
This cell holder holds one micro multi-cell, either 8 or 16 cells, for micro-volume
measurement. Two types of micro multi-cell holders are available: the standard
type (MMC-1600) and the constant-temperature water circulation type
(MMC-1600C).
8/16-Series Micro Multi-Cell Holder MMC-1600
8/16-Series Constant-Temperature Micro Multi-Cell Holder MMC-1600C
P/N
Model
206-23680-91
206-23690-91
Micro Multi-Cells
There are two types of micro multi-cells available for both the 8-series and
16-series models: a 50 L type and a 100 L type. The cell intervals of the
8-series micro multi-cells are applicable for use with 8 × 12-well microplates and
8-channel pipettes. Microplate samples aspirated into multi-channel pipettes
can be injected directly into the cells for measurement.
8-Series Micro Multi-Cell; optical path length: 10 mm; cell volume: 100 L
16-Series Micro Multi-Cell; optical path length: 10mm; cell volume: 100 L
8-Series Micro Multi-Cell; optical path length: 5 mm; cell volume: 50 L
16-Series Micro Multi-Cell; optical path length: 5mm; cell volume: 50 L
P/NModel
208-92089
208-92088
208-92086
208-92085
(Note)
Square cells (P/N 200-34442) are not included as standard. Please purchase separately.
(Note)
Square cells (P/N 200-34442) are not included as standard. Please purchase separately.