C101-E121B UV-VIS Spectrophotometers UV-2600/2700
Experience the Precision Desired, in Any Situation
Single monochromator UV-2600 Capable of a Measurement Wavelength up to 1400 nm Equipped with a single monochromator, providing low noise performance across a wide wavelength range Enables near-infrared measurements (up to 1400 nm)* * W he n t he opt i ona l I S R- 2600P l us i nt e gr a t i ng s phe re i s us e d Double monochromator UV-2700 Performance with a Minimum 8-Abs Photometric Range Equipped with an ultra-low stray light double monochromator, capable of 8-Abs measurements Uses the Shimadzu propr
Lo-Ray-Ligh grade diffraction grating Achieves Ultra-Low Stray Light, Enabling 8-Abs Measurements Absorbance Linearity UV-2700 Calibration Curve 9.0 8.0 6.0 Abs. Conventional models provide 5-Abs coverage. Even high-end models have been limited to 6 Abs. In contrast, the UV-2700 achieves ultra-low stray light levels, expanding the range to 8 Abs, with a transmittance value of 0.000001 % (1 part in 100 million). This system achieves high-level absorbance measurements with incomparable precision.
Double monochromator UV-2700 Performance with a Minimum 8-Abs Photometric Range Equipped with a double monochromator that achieves ultra-low stray light levels, the UV-2700 is optimal for measuring low transmittance samples, such as polarization films used for LCD panels. The UV-2700 is capable of 8-Abs measurements, and can make accurate transmittance measurements to 1 part in 100 million, accommodating a variety of sample measurements. Sample Polarization Film Measurement 6.
Single monochromator UV-2600 Capable of a Measurement Wavelength up to 1400 nm A key feature of the UV-2600 single monochromator type is its measurement wavelength range. By using the optional ISR-2600Plus Integrating Sphere attachment, the measurement wavelength range can be extended from 220 nm to 1400 nm, significantly expanding its applications.
Transmission Measurements of Polycrystalline Silicon Using the ISR-2600Plus UV-2600 Relative Diffuse Reflection Measurements of an Anti-Reflective Film Using the ISR-2600Plus 50.0 UV-2600 60.0 40.0 40.0 T% R% 30.0 20.0 20.0 10.0 0.0 500.0 0.0 600.0 800.0 1000.0 1200.0 1400.0 nm This is a transmission measurement of polycrystalline silicon. Since the system is capable of measurements to 1400 nm, the transmission characteristics of the band gap region (near 1000 nm) are clearly evident.
600 mm 28 % Size Reduction in Comparison to Conventional Models 120 mm 28 % Size Down Conventional model (UV-2450/2550) 60 mm 450 mm Compact 450 mm Width Size With the space-saving UV-2600/2700 models, the sample compartment size remains the same despite a 20 % reduction in installation width. A variety of film and other measurements can be performed effortlessly, without sacrificing user-friendliness.
Single monochromator UV-2600 Double monochromator UV-2700 Even More Compact and User-Friendly To respond to feedback that conventional models are too large, we have challenged ourselves to maximize space savings. The installation space required for this system has been reduced by about 28 %, thanks to a brand new compact design. In addition, validation software is provided as standard, so equipment inspections are easily performed, further enhancing user-friendliness.
Single monochromator UV-2600 Double monochromator UV-2700 With a Wealth of Accessories, Accommodates Any Application The functionality of the UV-2700/2600 can be freely expanded to suit the measurement objective. By accommodating a wealth of accsesories, the system can address any user’s applications and a variety of situations. Thanks to intuitive operations, anyone can easily obtain the data required.
Analysis Correspondence Chart UV-2600 UV-2700 Environment Electricity, Electronics, and Optics E F Quantitation of total phosphorus and total nitrogen E E in river water, lakes, and marshes E F Turbidity measurements E F Quantitation of iron, copper, arsenic, ammonia, and other substances in water F E E F E F E F E F Transmittance measurements for functional films Transmittance measurements for solar cell cover glass E F E F Construction Transmittance measurements for window gl
A Wealth of Applications Electricity, Electronics, Band Gap Measurements for and Optics Compound Semiconductors The diffuse reflection spectra for two types of compound semiconductors (red line: Culn0.5Ga0.5Se2, blue line: CulnSe2) used as solar cell materials have been measured using the ISR-2600Plus integrating sphere. It is evident that the absorption edge (position where the reflectance drops) differs depending on the sample. This difference signifies a difference in the band gap* for these samples.
Environment Hexavalent Chromium Measurements An aqueous solution of hexavalent chromium was colored with a diphenylcarbazide reagent, and the absorption spectrum was then checked. It is evident that the maximum absorption peak is at 542 nm. By using a calibration curve, the concentration of this sample was found to be 0.843 mg/L. Construction Window Glass Transmission Measurements Two types of window glass were measured utilizing the ISR-2600Plus integrating sphere.
UVProbe Software All-In-One Software UVProbe is an all-in-one software package equipped with the following four functions: Spectrum module Photometric module (quantitation) Kinetics module (time-course measurement) Report generator Each can be easily operated from its own special screen.
A Variety of Data Processing and Calculation Functions Data processing operations, such as peak detection and area calculation, and data conversion operations, such as differentiation and interpolation, can be applied to spectra and time-course data. With the kinetics module, the Michaelis constant (Km) or the maximum response speed (Vmax) can be calculated and plotted. Calculation Expressions and QA/QC Functions With the photometric module, calculation expressions can be defined for measurement results.
Optional Software UVProbe Agent Software (P/N 206-21550-92) The UVProbe Agent, which was developed for Shimadzu UV-VIS spectrophotometers, is used to automatically transfer and store the data acquired, or the results of performing data processing with the UVProbe software, to a general-purpose database, and to perform operations related to high-security data management and electronic signatures. This makes it possible for UVProbe to attain compliance with FDA 21 CFR Part 11.
Color Measurement Software (P/N 206-67449) This software calculates color values of the measurement sample from the spectrum measured.
Accessories (Options) Cells Description Optical Path (L) Required Sample Volume 10 mm Type 2.5 mL to 4.0 mL Fused Silica (S) (1) Glass (G) 200-34442 200-34565 200-34446 200-34446-01 200-34944 200-34944-01 200-34676 200-34676-01 200-34444 200-34444-01 20 mm 5.0 mL to 8.0 mL 50 mm 12.5 mL to 20.0 mL 100 mm 25.0 mL to 40.0 mL Square cell with stopper 10 mm 2.5 mL to 4.0 mL (2) Semi-micro cell 10 mm 1.0 mL to 1.
Universal Rectangular Cell Holder, Four-Cell Type (P/N 204-27208) Holds four rectangular cells with an optical path length of 10, 20, 30, 50, 70, or 100 mm. Reference-Side Rectangular Long-Path Absorption Cell Holder (P/N 204-28720) If using a 4-cell-type universal rectangular cell holder, use this as a referenceside cell holder if necessary. (Note) The Four-Cell Sample Compartment Unit (P/N 206-23670-91) is required.
Accessories (Options) 8/16-Series Micro Multi-Cell Cell Holders Model P/N 8/16-Series Micro Multi-Cell Holder MMC-1600 206-23680-91 8/16-Series Constant-Temperature Micro Multi-Cell Holder MMC-1600C 206-23690-91 This cell holder holds one micro multi-cell, either 8 or 16 cells, for micro-volume measurement. Two types of micro multi-cell holders are available: the standard type (MMC-1600) and the constant-temperature water circulation type (MMC-1600C).
Constant-Temperature Cell Holder (P/N 202-30858-04) Maintains a sample cell and reference cell at a desired, uniform temperature, by circulating constant-temperature water.
Accessories (Options) NTT-2200P Constant-Temperature Water Circulator (P/N 208-97263) Circulates temperature-controlled water to a constant-temperature cell holder. • Temperature range: Ambient + 15 °C to + 80 °C • Temperature control precision: ± 0.05 °C or more • Circulation pump: maximum flow rate 27/31 L/min; maximum lift 9.5/13 m (50/60 Hz) • External circulation nozzle: 10.5 mm O.D.
ASC-5 Auto Sample Changer (P/N 206-23810-**) Combine with a sipper unit or syringe sipper to build an automated multisample spectrophotometry system. • The aspirating nozzle is programmed to move in the X, Y, and Z (vertical) directions. • Up to eight sets of operational parameters, including the sizes of racks and the numbers of test tubes, may be stored in the battery backup protected files. • Up to 100 test tubes may be set together on the rack.
Accessories (Options) ISR-2600 Integrating Sphere Attachment ISR-2600Plus Integrating Sphere Attachment (P/N 206-28400-41) (for UV-2600 only) By combining the 0°/8° incidence angle integrating sphere with the S/R exchange function of the spectrophotometer, diffuse and specular reflectance measurements are possible without using any special attachments.
Absolute Reflectance Attachment Ref Model 100 % light P/N ASR-3105 Absolute Reflectance Attachment, 5° 206-16817 ASR-3112 Absolute Reflectance Attachment, 12° 206-16100 ASR-3130 Absolute Reflectance Attachment, 30° 206-15001 ASR-3145 Absolute Reflectance Attachment, 45° 206-15002 These accessories are intended for use with the MPC-2600 Sample Compartment, and require the BIS-3100 Sample Base Plate-Integrating Sphere Set (P/N 206-17059).
Hardware Specifications Item UV-2600 UV-2700 Double-beam optics Photometric system Photometric system Czerny-Turner mounting Single monochromator Lo-Ray-Ligh grade blazed holographic grating Czerny-Turner mounting Double monochromator Lo-Ray-Ligh grade blazed holographic grating R-928 Photomultiplier Detector 50 W halogen lamp, deuterium lamp, light source auto position adjustment built in Light source 185–1400 nm Setting wavelength range 185–900 nm Measurement wavelength range 220 nm to 1400 n
Software Specifications UVProbe Software Operating System Windows 7 Professional (64/32 bits) Data Acquisition Modes Spectrum, Photometric (quantitation), and Kinetics (time course) • Initialization results log • Wavelength accuracy • Multitasking (possible to execute data processing while measurement is being performed) General UV Validation Software • Wavelength repeatability • Photometric accuracy Inspection Items • Photometric repeatability • Resolution • Stray light • Customizable measurement
UV-2600/2700 Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its affiliates, whether or not they are used with trademark symbol “TM” or “®”. Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. For Research Use Only. Not for use in diagnostic procedures.