User's Guide

Table Of Contents
EVR-VPXDOC-0036 VP Series 2000 User Guide
DIVIGRAPH
Rev 2.4.1
September 2021
Proprietary & Confidential Copyright © 2020 Divigraph (Pty) Ltd. All Rights Reserved 15
8. To avoid damaging the e-module, you must first align the contact pins of
the e-module with the contact points in the case. Angle the e-module
and align the contact pins and back keyway.
9. Now rotate the e-module downward to align the notch on the other side
of the e-module with the notch in the case and press firmly. You can feel
or hear a click when it is in place. If necessary, hold the e-module in
place with your finger.
10. Inspect the exterior threads of the sensor case as well as the interior threads of the e-module
retaining ring to be sure they are not damaged.
11. Align the e-module retaining ring over the e-module. If needed, hold the
e-module in place with your finger.
12. To maintain the device’s IP rating, be careful to avoid damaging the O-rings. Turn the e-module
retaining ring clockwise until it seats against the O-ring before, fastening it with a reciprocating
motion (¼ turn back every turn) until the ring is hand-tight.
13. Verify that the device is properly sealed - there should not be any gap
between the e-module retaining ring and the case.