Specifications
Measurement modes of MOC63u
200°C
Ending modes
Set ending time
Ending drying
TemperatureTemperatureTemperature
TemperatureTemperature
Moisture
vaporization
Moisture
vaporization
Moisture
vaporization
Moisture
vaporization
Moisture
vaporization
Automatic ending
Moisture vaporization
over 30 seconds
Alternate drying modes
Starting mode
Choose the right measuring mode for your application.
30sec
5min
T3
T2
T1
t3t2t1
Automatic ending mode
Automatically ends measurement when
moisture loss over the previous 30
seconds becomes smaller than specified
percentage.
Timed ending mode
Automatically ends measurement when
the specified amount of time has
elapsed.
Rapid drying mode
First dries with the highest temperature
for the specified period, then shifts to
the specified temperature, shortening
measurement time.
Slow drying mode
Gently heats samples that might solidify
at the surface or samples that reduce
under high temperature.
Step drying mode
Allows step-by-step changes in drying
conditions. This feature is useful when
measuring samples that contain a large
amount of water.
Automatic starting mode
Starts measurement immediately after closing the lid. It will save time when conducting
repeated measurements.
Measurement modes
Timer setting
Interface
Measurement conditions
data memory
Data memory
Temperature calibration kit
Standard
(Easy start/Automatic end/Timed end)
Rapid drying
(Easy start/Automatic end/Timed end)
Slow drying
(Easy start/Automatic end/Timed end)
Step drying
(Easy start/Automatic end/Timed end)
1–120 minutes or continuous
(max 12 hours)
RS-232C (9-pin connector) I/O port
USB port
10
100
Option
Max
Min
60 g
0.02 g
0.001 g
0.01/0.1% (Selectable)
0.15% (2 g)
0.05% (5 g)
0.02% (10 g)
Straight type halogen heater
400 W
50–200°C (1°C increments)
5 to 40°C, 85%RH or lower
4 kg
(There is a time restriction when exceeding 180°C.)
LCD with backlight
ø95 mm
202
×
336
×
157
Capacity
Minimum readability
Repeatability
Drying heater
Power
Display
Pan size
Weight
Temperature range
setting
Operational temperature
and humidity range
Dimensions (W
×
D
×
H)
mm
Shimadzu Electronic Balances
General Catalog 23