Instruction manual
-26-
16.4 Command Code
Described in this section are the command codes which can be used
when your BL balance is connected to a computer via the IFB-102A
RS-232C interface.
The use of characters other than those described here will cause
errors in weighing and data transfer procedures. If an improper code
is mistakenly entered, disconnect the balance power cable for 10
seconds, then reconnect.
Command code Function Description
T
D05
D06
D01
D09
D07
D03
Q
{. }
Taring
Print (output once)
Autoprint
Continuous output
Output stop
Single output with stability
information
Continuous output with stability
information
ON/OFF selection
Echo back
Equivalent to the TARE key
Equivalent to the PRINT key
Refer to “16.1 The EP-60A Electronic Printer”.
Continuous output of data in the balance at
every approx. 100 ms.
For less than 1200 bps, it becomes approx. 150 ms.
Autoprint or continuous output is canceled.
Printing is made once with stability
information.
Continuous printing is made with stability
information.
Toggles between standby state and
measurement state.
Characters from these command codes until
delimiter are received and transmitted by every
character.
By use this command, message of personal
computer is able to print out EP-60A.
Characters length is under 16 characters,
including delimiter.
Caution