13. TROUBLESHOOTING
For countermeasures having an asterisk, contact the nearest Shimadzu sales or service
center.
When What trouble Cause -> Countermeasure
Before
weighing
Nothing is displayed by connecting
the AC adapter to the outlet.
is displayed.
The AC adapter is disconnected.
The electrical board is turned OFF.
There is an internal error in the balance. *
During
weighing
is displayed.
is displayed.
The display fluctuates.
The display does not change from
zero even if a sample having the
weight near readability is loaded.
The display slowly changes when
small amount of sample is loaded.
(Normal: readability/1 sec.)
has appeared suddenly.
Data communication cannot be
made.
The mass on the pan is too heavy.
Sensitivity is not correct.
The pan or the pan supporter is not in place.
Influence from vibration or wind
Improve the installation site.
Change the measurement mode to High-
stability mode.
Influence from electric noise or electromagnetic
wave
Maintain a proper distance from the noise
source.
Zero tracking works.
Refer to “6. Menu Selection”.
The averaging processing is in High-stability
mode.
Change the measurement mode if necessary.
There has been an instantaneous power failure.
Press the POWER/BRK key (the balance
enters weight display mode).
Setting of communication parameter is wrong.
Refer to “16.4 Setting the input/output format”.
Wiring of RS-232C cable is wrong.
During PCS
or % setting
has displayed. Set value exceeds the specified range.
Refer to “9. % setting” and “10. PCS setting”.
During
sensitivity
calibration
Does not proceed to next step.
(The stability mark does not light)
has displayed.
Influence from vibration or wind
Improve the installation site.
Change the measurement mode to High-
stability mode.
The weight used for sensitivity calibration is
wrong.
Check the weight and retry the sensitivity
calibration.
There is an internal error in the balance. *