Instruction manual

-17-
11. PERFORMANCE CHECKS
Conduct performance checks in a room where the temperature does not change suddenly.
These checks are used to determine if the balance conforms to specifications, and should be
performed with the greatest care.
Preparation Warm up the balance well.
Set the measurement condition as follows:
Repeatability Load and unload 10 successive times,
(1) an weight which is near the capacity of
the balance. Then record the following
items:
Xi: Displayed value when the weight
is loaded after stability mark is
lit.
Yi: Displayed value when the weight
is unloaded after stability mark is
lit.
(2) Calculate the standard deviation of
σ
x and
σ
y using the formulas shown
right.
(3) Balance operation is normal when
the standard deviation is less than
1.5 times the value specified.