Operation Manual

Dimension BST 1200 / SST 1200 User Guide
Setup
Workspace
Observe the following when placing Dimension in its operating
location:
Dimension has an approximate weight of 148 kg (326 lbs)
and requires a table capable of safely supporting 181kg
(400 lbs).
System Dimensions:
838 x 737 x 1143 mm (33 x 29 x 45 in)
Four-inch minimum space behind unit for air circulation
Dedicated outlet requirements:
110-120 VAC, 60 Hz, 15 A min. (20 amp recommended)
–or–
220-240 VAC, 50/60 Hz, 7 A min. (10 amp recommended)
Temperature: 18–30°C (65-86°F)
Relative Humidity: 30-70%, Non-condensing
Ethernet 10/100 Base T network
Optional UPS for power interruptions (brown-out
conditions):
Rated Power – 2200 VA
Output Power – 1600 watts