TROUBLESHOOTING
Behavior Problem Action
No lights on front panel Power switch off Turn on power switch on
rear of stand
Power not connected Insure cord is secure at
both ends and has supply
power
Power setting Verify supply power and
power module setting
Stand does not move Limit switch Attempt movement in other
direction.
Release limit switch.
Emergency stop Release emergency stop
button.
Stand moving at wrong
speed
Incorrect speed unit-of-
measure (in/min or
mm/min)
Press and together at
the same time to change
units of speed
Stand does not reverse at
limit
Manual mode selected Change mode to auto-
reverse
No reverse upon sample
break with GTX
Prior test peaks not cleared Press RESET on the GTX
prior to starting each test
GTX configuration Set break detect and verify
peak reading is larger
Gauge not in alignment
with fixture
Alignment adjustments See ‘Alignment’ section of
manual
Grinding noise
Non smooth movement
Struggling motor
Stand may have been
overloaded or other
problem
Contact your Dillon dealer
Console lights remain on
after power down.
Normal Allow 20 seconds for
capacitor to discharge and
lamps to extinguish.
Contact your Dillon dealer if these troubleshooting steps do not take care of the
condition.