DWP-98 Series Counting scale User’s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT Specifications subject to change without prior notice
CONTENTS 1. INSTALLATION 2. SPECIFICATIONS 3. KEYS, DISPLAY AND CONNECTIONS 4. GETTING STARTED 5. INITIAL SETUP 6. INSTRUCTION FOR USE 7. RS232 DATA OUTPUT 8. BATTERY POWER AND RECHARGING 9. ERROR CODES 10.
Appendix A: About Bi-Directional Communication INSTALLATION Because of metrological legislation, installation/some metrological parameter settings are limited to be done by authorized personnel only. Do not attempt to change any of the built-in metrological parameters. Contact your dealer for more information and technical assistance. To ensure performance accuracy, do not use the instrument in where or when the environment condition falls beyond as those listed on SPECIFICATIONS.
1. SPECIFICATIONS 2.1 GENERAL SPECIFICATIONS DWP-98 Series(resolution 1:60000; unit :kg/lb) Model# DWP-98CDH DWP-98CBH DWP-98CAH Max Capacity 3kg 15kg 30kg Readability 0.05g 0.2g 0.5g Interface Stabilisation time Operating temperature Power supply Display Zero range Housing ADC External Resolution Interface Other function Common Specification RS232 2 second typical 0°C - 40°C / 32°F - 104°F 220~240V AC (110V optional) 50/60Hz built-in rechargeable battery 6V4AH. 3 windows 6 digits 0.
2. KEYS, DISPLAY AND CONNECTIONS 1. UNIT PIECE WEIGHT SET KEY Press this key to confirm the unit piece weight entered by numeric keys. 2. WEIGHT UNIT KEY Press this key to shift among various weight units. 3. M+ KEY Press this key to accumulate the current quantity to memory or to recall the accumulated total quantity when unloaded. 4. TARE KEY Press this key to tare off the weight of a container. Press this key to enter parameter 5.
quantity check mode. Press this key as next parameter when setting 7. PRINT KEY Press this key to send print data to a printer. Press this key as last parameter when setting 8. ZERO KEY Press this key to set weight displayed to zero when an empty scale has drifted away from a true zero reading. 9. CE KEY Press this key to clear value entered. 10. NUMERIC KEYS Numeric keys 0 ~ 9 and decimal key. 11.
13. ZERO INDICATOR Visible when instrument at true zero weight status. 14. NET INDICATOR Visible when tare function is in effect. The weight being shown on Weight Panel is net weight. 15. STABLE INDICATOR Visible when weight reading is stable. 16. WEIGHT UNIT INDICATOR kg = kilogram, g = gram, lb = pound 17. UNIT PIECE WEIGHT INDICATOR The value left to it is the unit piece weight. This indicator remains visible during normal operation status. 18.
A. WEIGHT PANEL Weight value is shown here. B. UNIT PIECE WEIGHT PANEL Unit piece weight is shown here. C. QUANTITY PANEL Quantity value is shown here. D. BUBBLE LEVEL Refer to this bubble to obtain a horizontal position for the instrument. E. CHARGE STATUS INDICATOR Red = Recharging battery; Green = Charging completed. F. ON/OFF KEY Press this key to turn the instrument on or off. G. DC JACK INPUT External power adaptor is plugged in here. H.
4. GETTING STARTED In order to obtain accurate weighing and counting result, the instrument must be placed on a strong and level surface horizontally. Avoid using the instrument in environment where excessive wind flow, vibration and extreme temperature change exist. The instrument should be installed CA from any sources of excessive electrical noise. For full EMC or for RFI immunity, termination of cable shields and correct grounding of the instrument is essential.
4.3.1. Connecting with a Computer RS232 COMPORT ON INSTRUMENT COM PORT ON COMPUTER (DB9) (DB25) 2 = RXD 3 = TXD 3 = TXD 3 = TXD 2 = RXD 2 = RXD 5 = GND 5 = GND 7 = GND 4.3.2 Connecting RS232 to Printer (DB25) RS232 COMPORT DB25 COMPORT ON ON INSTRUMENT PRINTER 2 = RXD 3 = TXD 3 = TXD 2 = RXD 5 = GND 7 = GND 5. INITIAL SETUP 5.1 INTERNAL SETTINGS Preferred application parameters can be checked or set through internal functions. Refer to 5.
c. The instrument displays F1, d. The instrument is now in internal function, 5.
6. INSTRUCTION FOR USE 6.1 POWER ON Powered on the instrument, instrument displays: a. softCAre and revision number and capacity of instrument, b. calibration count value, c. parameter set count value, d. all LCD segments, Then countdown process starts. After that, the instrument is ready for operation. 6.2 START WEIGHING a. If zero weight cannot be obtained when unloaded, press [ZERO]. After [ZERO] is pressed, Zero Indicator appears. Refer to SPECIFICATIONS for maximum zero range, b.
6.4 TARE OFF THE WEIGHT OF A CONTAINER 1 Tare function is used to temporarily set the instrument to zero (such as cancelling the weight of a box or a container) in order to get the net weight result 6.4.1 Enable / Disable Repeat (Multiple) Tare Depends on internal function setting, repeated (multiple) tare operation may be enable (Mode 1) or disabled (Mode 2). Contact your dealer for more information. When repeat (multiple) tare is enabled: 1.
f. To cancel tare effect, remove all loads and container from platter and press [TARE] , g. Net Indicator disappears. Tare effect has been removed and weight reading is gross weight. 6.4.3 Preset Tare This feature enables the user to manually enter the tare weight. During normal operation, enter the pre-determined tare weight through numeric keys then press [TARE]. This pre-determined tare value will be deducted. To cancel the preset tare effect, remove all loads from platter then press [TARE]. NOTE: 1.
d. To enter new unit piece weight, repeat step b above. 6.5.2 Sampling Process This method is used where unit piece weight is unknown or the weight of individual piece is not highly standardized. a. Refer to 6.2 and 6.4 for zero, selecting preferred weight unit and tare operation, b. Place a sample size with known quantity on platter c. Enter the quantity of the sample through the numeric keypad and confirm by pressing [# SET]. d.
b. The quantity added to platter is less than 100% of the maximum counts previously attained from the same transaction. If the above requirements are fulfilled, a new unit piece weight will be determined and displayed on the Unit Piece Weight Panel and confirmed by 2 audio beeps. Auto Unit Piece Weight Enhancement Function will terminate when: a. a zero weight is detected or during the transaction process, or b.
10 c. output transaction printout(s) through the RS232 comport. The instrument returns to normal operation status after 2 seconds, 11 Repeat a for subsequent transactions , 6.7.2 Memory Recall and Clearance a. Remove all loads from platter, b. Press [M+] to recall total accumulated weight from memory, c. After [M+] is pressed, the instrument displays: d. The instrument displays: “ACC” on Weight Panel. “n” on Unit Piece Weight Panel.
Hints: 1. For normal comparison, set both HI and LO limits, 2. To check only if result is lower or equal to LO (result ≤ LO), set HI limit = 0, 3. To check only if result is higher or equal to HI (result ≥ HI), set LO limit = 0, 4. To check if result is equal to a specified value, set both HI limit and LO limit = the specified value 6.8.2 Visual and Audio Quantity Check Signals 6.8.2.1 Visual signal: - Quantity check result is displayed by one of the Check Result Indicators located on Quantity Panel. 6.8.
7. RS232 DATA OUTPUT Data output parameters are fixed as below: Data Bit = 8 Parity = None Stop Bit = 1 13 There are 2 data output modes (PC and Prt) are available . PC is for communication with computer and other peripherals which accepts and processes continuous data communication. Prt (printer) is for transmission to printer or other peripherals which accept only single or manual data transmission. Baud rate has to be set before proceeding to other settings. 7.
space. 14 15~16 COMMA SEPARATION UNIT kg = kilogram g_= gram (_ = blank space) lb = pound 17 Cr 18 LF 7.2 PRT (PRINTER) MODE 7.2.1 Printing Current Transaction A standard printout as will be sent through the RS232 Comport when 14 [PRINT] is pressed. Refer to Standard Printout Sample on next page for description. Number of copies sent = the number of copy set in internal function F8. Refer to 5 INITIAL SETUP for more information. If extra copies are needed, press [PRINT] again. 7.2.
Standard Printout Sample TIME 12:08:57 DATE 07.08.2010 No. 0 TARE 2.000kg GROSS 6.000kg NET 4.000kg UNIT WT. 0.5 g Q'ty PCS HIGH LOW HIGHER 5000PCS 4500PCS THAN LIMIT Description 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. Time of printing, Date of printing, 16 M+ sequent number Tare Weight, Gross Weight, Net weight, Unit piece weight, Quantity, One blank line, One blank line, ** HI limit set ** LO limit set ** Quantity Check result Totalized Accumulated Printout Sample TIME DATE No.
8.1 SYMBOLS AND REMAINING POWER: Full Battery: ≥ 6.3V 2 Blocks: ≥6.0V 1Block: ≥5.7V Frame only: <5.7V 8.2 BATTERY OPERATION TIME Depends on the battery condition, a fully charge new rechargeable battery will usually give: Around 200 hours of continuous operation without backlight, or Around 70 hours of continuous operation with backlight on, or 8.3 RECHARGE BATTERY When appears, (when battery is less than 5.7V), it means the built-in rechargeable battery is at low voltage status.
Err 1 Time value error Err 2 Date value error Err 3 Exceed maximum manual zero range Err 4 Offset out of range / unstable during power on Err 5 No load cell signal detected Err 6 Tare operation error Err 7 Err 8 Logic error. HI limit set is lower than LO limit (and HI is not = 0) Logic error. LO limit is higher than HI limit (and HI is not = 0) --oL-- Overload (Gross weight is more than Max plus 9d) UndEr Under load (Gross weight is less than minus 20d) 10.
About Bi-Directional Communication Computer commands can be sent to this instrument through any standard communication program like Hyper Terminal which comes with Windows 2000 and Windows XP. Equivalent commands from computer are illustrated on below table for each of the key on panel. 1 Equivalent to Computer Command 1 2 2 3 3 4 4 5 5 6 6 7 7 8 8 9 9 0 . (Decimal) ZERO 0 .
Calibration Turn on the scale and press/hold the tare key for 4 sec then Let go and it will display F1, with the “Print” button go to F10 Cal Press “Tare “ to enter ,show “P----”; the initial pass ward is 1234 . (“Alarm” for next parameter; ”Print” for last parameter ) Press “Tare “ to enter to “P1 Cap” (capacity : 3kg,5kg,6kg,10kg,15kg,20kg,30kg) “”P2 Inc” (1,2) “P3 Zero track ” (off, 0.25, 0.5, 1, 1.5 , 2.