DWP-102E Weighing Indicator User’s Guide
Table of Contents SECTION 1 INTRODUCTIONS………………………………………………………………1 1.1 Introduction……………………………………………………………………………………1 1.2 Specification…………………………………………………………………………………..1 1.3 Installation…………………………………………………………………………..………...1 SECTION 2 KEYS DESCRIPTIONS………………………………………………………2 SECTION 3 OPERATIONS…………………….……………………………………….…..3 3.1 Power on/off................................................................................... 3 3.2 Functions and parameters setting...................................................
-1SECTION 1 INTRODUCTIONS 1.1 Introduction The indicator provides an accurate, fast and versatile series of general purpose weighing indicator with counting, check-weighing, animal weighing functions and preset tare. 1.2 Specification plastic housing Housing Power supply 110V 50/60Hz built-in rechargeable battery. 6V4AH. Display 6 digits 1’LED.
-2Power on/off the indicator. ZERO Key 1) When self-checking, Press ZERO Key to enter parameter setting; 2) Set the zero point for all-subsequent weighing; 3) Move the cursor to change the digit when setting parameters or other functions. TARE Key 1) When self-checking, Press TARE Key to enter fast calibration; 2) Tare the scale. Store the current weight in memory as a tare value, subtract the tare value from the weight and shows the net weight.
-3SECTION 3 OPERATIONS 3.1 Power on/off Press ON/OFF Key to power on the Indicator and the indicator self-checks, then goes into the weighing state. 3.2 Functions and parameters setting When self-checking, press ZERO Key to enter parameter setting. Then display shows “F1 CLK”. Press TARE Key to move functions cycle. Press FUN Key to enter the desire function or MOD Key to cancel or quite. See details on Section 4 Functions and Parameter Sections.
-4The scale should be tare a second time if another is added to the first one. Again only the weight added after tare will be displayed. When the container is removed a negative value will be shown. If the scale was tare just before removing the container this value is the gross weight of the container plus all products that were removed. The ZERO indicator will also be on because the platform is back to the same condition it was when the ZERO Key was last pressed.
-5weights, and then total weight. Clear the accumulated weight from memory: Press ZERO Key during display of the nos. of weights and total weight, accumulated weights are cleared from memory. 3.7 Holding the weight (KEY=3) Press MOD Key to enter the hold function. After every weighing, the weight value will be kept on the display. MOD Key can be pressed to cancel hold function. 3.8 Presetting tare value (KEY=4) Press MOD Key to preset tare value.
-6indicator HI will be on, while lower than the low-limit indicator LO will be on. Otherwise indicator HI & LO will be both on. Mode 2: BEEP=2, beep when it is eligible; Mode 3: BEEP=3, beep when it is not eligible. Note: The weight must be greater than 9 scale divisions for the check weighing to operate. To disable the check-weighing function press UNIT Key to set both high and low limit to zero. 3.11 Print (KEY=2) Set “P10 PS”: “RS Prt” is selected for the Serial Printer.
-7KEY 5 To set the MOD Key for animal weighing parameters setting. “F4 CHK”: Check-weighing setting Options: BEEP 1/2/3. “F5 Brt”: Adjust the brightness of LED display. For LED version Select 1-8 : 8 is most bright “F5 INP”: Display the internal counts For LCD version . Used to check the connection of loadcell and working status of ADC circuit. Press “TARE” to exit. “F6 INP”: Display the internal counts For LED version “F6 CAL”: Calibration for LCD Version.
-8“P12 KB”: When “on”, keyboard is pressed with sound. “P13 UT”: Unit setting. Lb, U1 or U2 is optional, which could be “ON” or “OFF” through pressing the FUN Key and TARE Key. When U1/U2 is “ON”, the conversion coefficient can be set through pressing UNIT Key while self-checking. When U1/U2 is “OFF”, it could not be set unless entering the P13 UT menu and set it “ON”. “P14 Pn”: Press “FUN” to display “P1 ----” set the passwords and press “FUN”. It display “P2 ----” and set passwords again to confirm.
-9SECTION 6 ERROR CODES ERROR DESCRIPTION RESOLUTION - - OL - - Over range Remove weight from the scale. Err 1/Err 2 Date and Time not right Reset again. Err 4 Zero Setting Error The scale was outside the normal zero setting CODES range either when it was turned on or when the ZERO Key was pressed. Remove weight from the scale and try again. Use the TARE Key to set the display to zero value.