Service manual

8508A
Service Manual
1-8
Maximum Voltage and Current Inputs
Notes to maximum voltage and current input specifications
Maximum DC input equal to maximum rms input. Maximum peak input is rms x 1.414
Specifications apply equally to front and rear input terminals except where noted below.
Front to rear isolation allows opposing polarity of maximum terminal voltage on each input.
Digital I/O Ground (DigGnd) is internally connected to Safety Ground (Ground).
DC and AC Voltage
Maximum rms terminal voltages
Sense Hi
Hi 250 V
Sense Lo 1000 V 1000 V
Lo 250 V 1000 V 1000 V
A 1000 V 1000 V 1000 V 1000 V
Guard 1000 V 250 V 250 V 1000 V 1000 V
DigGnd 650 V 650 V 650 V 650 V 1000 V 1000 V
Ground 0 V 650 V 650 V 650 V 650 V 1000 V 1000 V
Notes
The A terminal is open circuit in these functions.
In 4wV mode Sense Hi is internally connected to Hi and Sense Lo is internally connected to Lo.
DC and AC Current
Maximum rms terminal voltages
Sense Hi
Hi 250 V
Sense Lo 1000 V 1000 V
Lo 250 V 1000 V 1000 V
A 5 V 250 V 1000 V 1000 V
Guard 250 V 250 V 250 V 1000 V 1000 V
DigGnd 650 V 650 V 650 V 650 V 1000 V 1000 V
Ground 0 V 650 V 650 V 650 V 650 V 1000 V 1000 V
Maximum rms terminal currents
Guard A Lo Sense Lo Hi Sense Hi
Front Input n.a 20 A 20 A n.a n.a n.a
Rear input n.a 2 A 2 A n.a n.a n.a
Notes
The Sense Lo, Sense Hi, and Hi Terminals are open circuit in these functions.
The front input A terminal protection is automatic and self-resetting, and does not interrupt current flow. Damage is likely to
occur if more than 20 A is applied.
The rear input A terminal is protected by a fuse mounted on the rear panel.
Resistance and Temperature
Maximum rms terminal voltages
Sense Hi
Hi 250 V
Sense Lo 250 V 250 V
Lo 250 V 250 V 250 V
A 250 V 250 V 250 V 250 V
Guard 250 V 250 V 250 V 250 V 250 V
DigGnd 650 V 650 V 650 V 650 V 1000 V 1000 V
Ground 0 V 650 V 650 V 650 V 650 V 1000 V 1000 V
Note
The A terminal is open circuit in these functions.
riešenia na presné meranie
http://www.elso.sk