Service manual

Maintenance
Self Test 4
4-15
AC VOLTS
Test No 80 Err 2801 P015 100mV AC Range: Zero Input Check
Test No 81 Err 2811 P016 100mV AC Range: +179mV DC Input Checks at
PreAmp Output
Test No 82 Err 2821 P017 100mV AC Range: -179mV DC Input Checks at
PreAmp Output
Test No 83 Err 2831
2832 raw
P018 1V AC Range: Zero Input Check
Test No 84 Err 2841 P019 1V AC Range: +993mV DC Input Checks at
PreAmp Output
Test No 85 Err 2851 P020 1V AC Range: -993mV DC Input Checks at PreAmp
Output
Test No 86 Err 2861 P021 1V AC Range: +993mV DC Input Checks at RMS
Converter Output
Test No 87 Err 2871 P022 1V AC Range: -993mV DC Input Checks at RMS
Converter Output
Test No 88 Err 2881
2882 raw
P023 10V AC Range: Zero Input Check
Test No 89 Err 2891 P024 10V AC Range: +9.78V DC Input Checks at
PreAmp Output
Test No 90 Err 2901 P025 10V AC Range: -9.78V DC Input Checks at PreAmp
Output
Test No 91 Err 2911
2912 raw
P026 100V AC Range: Zero Input Check
Test No 92 Err 2921 P027 100V AC Range: +9.78V DC Input Checks at
PreAmp Output
Test No 93 Err 2931 P028 100V AC Range: -9.78V DC Input Checks at
PreAmp Output
Test No 94 Err 2941
2942 raw
P029 1000V AC Range: Zero Input Check
Test No 95 Err 2951 P030 1000V AC Range: +9.78V DC Input Checks at
PreAmp Output
Test No 96 Err 2961 P031 1000V AC Range: -9.78V DC Input Checks at
PreAmp Output
riešenia na presné meranie