User Manual
Table Of Contents
- 1. What’s in the Box
- 2. Camera Connection
- 3. Installation Diagram and Considerations
- 3.1. Installation Diagram
- 3.2. Installation Considerations
- 4. Monitoring PC System Requirements
- 4.1. Hardware
- 5. Installation Environment Before Installation
- 5.1. Check Windows 10 Version
- 5.2. Windows Update
- 5.3. Display Setting
- 5.4. Power and Sleep
- 6. Software Download and Installation
- 6.1. Download Software and Camera Network Configuration
- 6.2. Install the Software
- 7. Program Installation
- 8. Software Elements
- 8.1. Connection Information
- 8.2. Functions
- 8.3. Maximum Temperature Indication Window
- 8.4. Alarm setting
- 8.5. History
- 8.6. Top Navigation Bar
- 8.7. Main Display
- 8.8. Blackbody Status
- 8.9. Status
- 9. Run Program
- 10. Specifications
- 11. FAQs
- 11.1. Usability
- 11.2. Hardware Related
- 11.3. Compatibility
- 11.4. Compliance
- 12. Warranty Information
- 13. Limits and Exclusions
DW E.S.T. (Elevated Skin Temperature) System
-5-
3.2. Installation Considerations
When considering where to place your screening station, consider the following:
• Avoid placing your camera, Blackbody or people being screened directly under an AC vent or
a strong fluorescent light. The air temperature and heat emitted from the light may impact the
accuracy of the temperature readings.
• Do not place the camera facing a window or large door. These may reflect strong light and
heat from outdoors and have an impact on the accuracy of the temperature readings.
• Consider that any extensive physical activity performed before screening may impact the facial