Instruction manual

MXS-EVO - MANUAL
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
MXS-EVO – Manual – Rev 1.7 - 32 -
Custom Settings
The device permits to have some ad-hoc customizations through the unlock with particular activation codes.
Pulse Generation
The device is able to generate a programmable pulse in place of the IRIG-B DCLS signal (on the BNC connector) or the
PPS signal.
The following parameters can be set:
- <Start Time> HH:MM:SS
- <Stop Time> HH:MM:SS
- <Repetition Rate> MM:SS
- <Pulse Length> SS.ssss
The system generates a sequence of pulses of <Pulse Length> length and repeating
every <Repetition Rate> starting at <Start Time> and stopping at <Stop Time>.
Please note that the <Start Time> is always synchronous to the internal PPS
(eventually disciplined via an external source).
Clock Generation
The device is able to generate two distinct programmable square wave clocks by dividing two different sources with
integer dividers:
10 MHz
16.384 MHz