Specifications

connection of wires).
3. Remove clip holding thermocouple in place (1
screw) and remove the two screws on the ther-
mocouple terminals.
4. Remove the thermocouple. Pull thermocouple
straight out of hole in the chamber first to avoid
damage to the insulation.
5. Guide the looped ends of the new thermocouple
through the plastic bushings with the red (-) lead
to the right as you face the back of the furnace.
6. Form a 90° bend, towards the furnace, between
the 2" long and 1” long ceramic insulators with
the red (-) lead to the right. Insert the thermo-
couple straight through the hole in the chamber.
7. Secure the thermocouple with clip and screw.
Connect the looped ends of the thermocouple to
the terminals with + to + (yellow wire).
Chromel/Alumel thermocouples and lead wire
are easily tested with a magnet. The non-mag-
netic wire is positive (+) and the magnetic wire
is negative (-).
8. Replace both back covers.
9. Reconnect to power supply. Test operation of
furnace.
To Replace Solid State Relay
1. Disconnect furnace from power supply.
2. Remove back control cover. (Note connection
and placement of wires to relay).
NOTE: For basic automatic control models,
remove both the back cover and the front con-
trol panel to provide access to solid state relay.
3. Disconnect wires from terminals. Identify or
mark wires.
40
MAINTENANCE AND SERVICING