Specifications

Furnace Programming
Refer to “Programmable Models” in this manual to pro-
gram in the following values for performing ASTM specifi-
cation D3174.
PR1 (Ramp 1) value is 8°C/minutes
PL1 (Level 1) value is 500°C
PD1 (Dwell 1) value is 0 minutes
PR2 (Ramp 2) value is 6°C/minutes
PL2 (Level 2) value is 750°C
PD2 (Dwell 2) value is 120 minutes
PR3 (Ramp 3) value is END (for 8 ramp & 8
dwell controller)
PLC (Loop Center) value is 1
Included with this furnace are two stainless steel trays
which will hold crucibles of quartz jars. Use the appropri-
ate side for the type of crucible you are using. A remov-
able handle is also provided to use in loading and unload-
ing the trays from the furnace.
Shelf Location
1. The side walls of the chamber contain twenty
four holes (.250 inch dia.) which allow adjust-
ment of the two perforated shelves at various
heights within the chamber.
2. Using the eight, .250 inch diameter porcelain
pegs, insert four pegs in the bottom row of
holes.
3. Next insert the remaining four pegs in fourth row
of holes and insert the two perforated shelves .
4. Check for proper fit at this point by inserting the
crucible trays.
Operation
1. Insert crucible trays. One tray per shelf.
2. Close the furnace door.
34