Specifications

a few times, note that the amber furnace cycle light will
go out while the door is open. If the furnace cycle light
does not go out while the door is open, consult the
Troubleshooting section before proceeding.
Parameters
Pnr - Program Number. The program number of the pro-
gram you are going to work with. By pushing the up or
down button you can select a program numbered from 1
to 4.
PR1 - Program Ramp Rate. The rate of heat increase or
decrease in °C/minutes. Pushing the up or down button
will give current setting of this ramp.
PL1 - Program Level. The temperature to which the fur-
nace needs to attain. Push up or down button to set.
PD1 - Program Dwell 1. Amount of time in minutes to
hold PL1 program level temperature entered. Push up or
down button to set.
You will use the same descriptions and procedures used
for PR1, PL1, PD1 for the remaining Program Ramp
Rates PR2 - PR8, Program Levels PL2 - PL8, and Pro-
gram Dwells PD2 - PD8.
Cnt - Continue. Allows linking of programs. You may se-
lect Cnt as “y” (yes) or “n” (no) by pushing the up or down
button.
HB - Holdback. Automatically places the programmer
into “Hold” if the measured value deviates more than a
specified amount from programmer set point. When mea-
sured value re-enters the holdback band, the timing for
the segment resumes. (Parameter is expressed in °C and
only functions when running a program). Push up or down
button to set.
PLC - Program Loop Count. The number of times a pro-
gram will be repeated. Push up or down button to set.
SP1 - Set point One. Indicates current set point. Push up
or down button to set.
25
MULTI-PROGRAMMABLE MODELS
Note
The two center push buttons are inac-
tive and not used.