Specifications
the enter button again to set the new value as the current
value. (See “Programming Controller” for information on
programming.)
The control parameters in order are:
LoCt - Parameter lock level. Determines the number of
parameters displayed when you press “SEL”. It should be
set to “0002.” If it does not, set to “0002.”
Nod - Mode. By pushing “UP” or “DOWN” buttons, this
can be set to “Auto” or “NAn” (Manual). Set to “Auto”. In
Manual mode, you must enter values to control the fur-
nace chamber temperature.
At - Autotuning feature. Will automatically set optimum
P.I.D. values for your process. For Tuning, set to “on.” For
normal operation, set to “oFF.”
StAt - Status. Displays current status of program and
cannot be changed.
Tine - Time. Displays the time remaining in the current
program.
ProG - Program options. Can be set to one of three op-
tions: “oFF”, “rUn” or “hold”. “oFF” will stop any running
program and reset the program to the beginning. “rUn”
will start a program from the beginning or resume a pro-
gram from “hold” from the point it was held. “hold” will
interrupt a running program without resetting the program
to the beginning.
Sü-1 - Set point 1. The temperature at which the first
ramp is aiming and at which the first dwell will be held.
tN1r - Ramp time 1. The time the controller will take to
charge the chamber temperature from its present temper-
ature to the value of Set point 1. This time may be limited
by the physical capabilities of the furnace.
tN1S - Dwell time 1. The length of time the controller will
hold the furnace at the Set point 1 temperature.
Sü-2 - Set point 2.
tN2r - Ramp time 2.
19
PROGRAMMABLE CONTROL MODELS