Specifications

Changing Temperature
Indication
Push “SCROLL” button once, “°C” will appear. This indi-
cates temperature measurement. (Contact Thermolyne if
control needs to be changed to °F.)
Setting the High Alarm
Temperature Over Temperature
Protection (OTP)
Push “SCROLL” button until “AL.SP” (Alarm Set point) ap-
pears. The lower display should indicate 1100°C.
Depress either the “UP” or “DOWN” button to select the
OTP value you desire. Thermolyne recommends that you
set the value either at the maximum operating tempera-
ture of the furnace or a value of 20 degrees above your
working temperature if you desire to provide protection for
your workload.
17
AUTOMATIC MODELS
Note
HIAL-HIGH ALARM (over temperature
protection - OTP). The controller is fit-
ted with a mechanical relay which is
de-energized in the alarm mode. This
relay, when de-energized, removes
power from the heating elements. If
the primary control circuit fails, the
OTP will control the furnace tempera-
ture at the preset value you have
entered. It does not shut off the fur-
nace but will maintain the chamber
temperature at that value.