User's Manual
Table Of Contents
- Operator’s Manual
- Digital
- Control
- Incorporated
- DCI Headquarters 19625 62nd Ave. S., Suite B-103 Kent, Washington 98032 USA Tel 425 251 0559 / 800 288 3610 Fax 253 395 2800 E-mail DCI@digital-control.com www.digitrak.com
- Important Notice
- Safety Precautions and Warnings
- Safety Precautions and Warnings (Continued)
- _
- Introduction
- _
- Receiver
- Power On
- Toggle and Trigger Switches
- Speaker and Audible Tones
- Adjusting Screen Contrast
- Main Menu
- Locate Menu
- Set US Menu
- Low Fre/High Fre Menu
- Configure Menu
- Changing the Telemetry Channel
- 1-Point Calibration
- 2-Point Calibration (In-ground Calibration)
- Changing the Grade Mode
- Changing the Depth Measurement Mode
- Cold Screen / Normal Screen
- Tele Option A/B
- Locator DL / No Locator DL (Enabling and Disabling DataLog Menus)
- Set Roll / Unset Roll (Enabling and Disabling Roll Offset Function)
- Remote Display
- _
- Transmitter
- _
- Battery Charger
- _
- Locating
- _
- The Target Steering® Function
- _
- Cable System
- Cable System Components
- Non-DCI Supplies Required for Operating the Cable System
- Connecting Power Supply to Power Source and Cable Transmitter
- Grounding the Cable Transmitter
- Cable Transmitter On/Off
- Calibrating the Cable Transmitter
- Enabling the Roll Offset Function on the Remote Display
- Locating Using the Cable System
- Viewing Transmitter Depth or Predicted Depth
- Viewing Status of Cable System Power Source
- Target Steering Function Using the Cable System
- Troubleshooting
- _
- Appendix
- Depth Increase in Inches (Centimeters) per 6-foot (1.8 meter) Rod
- Depth Increase in Inches (Centimeters) per 10-foot (3-meter) Rod
- Depth Increase in Inches (Centimeters) per 15-foot (4.6-meter) Rod
- Percent of Grade to Degree Conversions (0.1% Pitch Transmitters or Sensitive Pitch)
- Degree to Percent of Grade Conversions (0.1% Pitch Transmitters)
- Calculating Depth Based on Distance Between FLP and RLP
- LIMITED WARRANTY
Receiver
6. Return to the main menu screen by pushing the toggle down one time.
7. Select Configure on the main menu display, and click the trigger.
8. Select 1 Pt. Cal., and click the trigger.
9. Click the trigger (arrows will have already selected High Fre Cal).
10. Toggle right to select Y for yes, and click the trigger.
11. Follow the instructions on the display, and click the trigger appropriately.
12. Return to the main menu by pushing the toggle down two times.
13. Select Locate, and click the trigger.
14. Place the receiver at 10 ft (3 m) and verify that it reads 10 ft (3 m) by holding in the trigger. Check the
depth readings in two other locations (e.g., 5 ft/1.5 m and 15 ft/4.6 m).
2-Point Calibration (In-ground Calibration)
NOTE: In-ground calibration is rarely needed. If you must calibrate with the transmitter in the
ground, use this procedure with caution.
2-Point Calibration for Standard, Mini, and Long-Range Transmitters
To calibrate the receiver using the 2-point calibration method with the standard, mini, or long-range trans-
mitter in the ground:
1. Verify that you see Low Fre on the main menu screen. If you see High Fre, then select it and click
the trigger so that the menu option will change to Low Fre.
2. Select Configure, and click the trigger.
3. Select 2 Pt. Cal., and click the trigger.
4. Select High Fre Cal, and click the trigger.
5. Toggle to select Y for yes, and click the trigger.
6. Position and stabilize the receiver at least 6 in. (152 mm) above the ground directly over the
transmitter; be sure that the locate line (LL) is aligned with the horizontal cross hairs to ensure you
are directly over the transmitter (refer to the Locating section for details on the LL).
7. Click the trigger when the display instructions indicate.
8. Raise the receiver at least 30 in. (762 mm) and stabilize it, then click the trigger.
9. To exit the calibration function and return to the main menu, push the toggle down two times.
DigiTrak
®
Eclipse
®
Operator’s Manual 23