Data Sheet

2/10/2018 OpenScope MZ Reference Manual [Reference.Digilentinc]
https://reference.digilentinc.com/reference/instrumentation/openscope-mz/reference-manual 18/20
Any of the 10 LA signal channels in any combination
Pattern matching of the LA signal channels is not supported
8 out of 8 DMA channels available on the PIC32MZ are used on the OpenScope MZ. Two channels are dedicated to UART, a pair of
DMA channels are assigned to each interleaved ADC () channel for a total of four channels, one channel is shared between the function
generator and the logic analyzer and one dedicated to hardware protocol communication. All DMA channels can be triggered on any
interrupt event and do not require the use of an ISR.
DMA is used to transfer data without utilizing the CPU by working in parallel with the CPU and has to ability to access peripherals and non-
cached memory at much greater speeds. DMA cell transfers are serialized so care is taken in the OpenScope firmware to ensure that
multiple channels are not triggered above 10 MT/s to prevent a high priority channel from stalling all other DMA channels.
9 PWM channels of the PIC32MZ are implemented on the OpenScope MZ. Two are used in the interleaving of the ADC () channels, three
are used for offsets for the AWG and both analog input channels, two are used for the DC outputs, and the remaining two used as offsets
for the DC outputs.
The PWM outputs have 330 unique values ranging from 0 V to 3.3 V, with a step size of 10 mV. The internal clock runs at 100 MHz (),
providing a PWM frequency of 303 kHz (). All of the analog designs on the OpenScope MZ are based on PWM values from 50 to 300 to
allow some headroom for calibration.
The LEDs on the OpenScope MZ are used to indicate the current status of the OpenScope MZ hardware as follows:
Note: Firmware versions prior to 1.2.0 will not exhibit the LED () behavior described below.
Blue Off - Device is booting and not ready to use.
Blue Flashing - Device is booted and ready to use but Wifi is not connected.
Blue Solid - Device is booted and ready to use and Wifi is connected.
The three other LEDs blink the last octet of the OpenScope MZ's IP Address.
Red Solid - Calibration or acquisition in progress.
All user LEDs Solid - An error has occurred. Reboot the OpenScope MZ.
When connected to a Wifi network the 3 user LEDs display the last octet of the OpenScope MZ's IP address by blinking the number
of times corresponding to that digit of the last octet in decimal. For example an OpenScope MZ with an IP address ending in '123'
would blink LD1 once, LD2 twice and LD3 three times.