Datasheet

The Analog Discovery Studio can be used with WaveForms' Oscilloscope instrument to capture analog input data via the
analog input (Scope”) channels, using either BNC cables or MTE cables. When this instrument is used, the Analog
Discovery Studio's analog input channels act as a two-channel, 14-bit, 100 MS/s oscilloscope.
When used with BNC cables, the oscilloscope channels are single-ended, however, the circuit under test must still share a
common ground with the Analog Discovery Studio. With BNC cables, the analog input channels have a bandwidth of
30+ MHz.
When used with MTE cables, the oscilloscope channels' + and - pins are differentially paired. The - pins can be attached to
a non-ground circuit net, but the circuit under test must still share a common ground with the Analog Discovery Studio.
With MTE cables, the analog input channels have a bandwidth of 9 MHz.
Since the Analog Discovery Studio's analog input channels are shared, the Oscilloscope instrument cannot be used at the
same time as the Voltmeter, Data Logger, Spectrum Analyzer, Network Analyzer, or Impedance Analyzer instruments.
For more information on the analog input (Scope”) channels, please visit the Analog Discovery Studio Specifications.
For a walkthrough of the different features of WaveForms' Oscilloscope instrument, please visit the Using the
Oscilloscope guide.
Important Note: Grounding Circuitry
Waveform Generator