IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 2 of 24 Contents: 1. 2. 3. 4. 5. 6. 7. 7.1 7.2 7.3 7.4 7.5 8. 9. 10. 11. 12. 13. 14. 15. Overview............................................................................................................................. 3 Models ..........................................................................................
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document Type Page Document title 1. Instruction Manual 3 of 24 Overview The HygroClip DI (Digital Interface) allows connecting up to two HygroClip digital humiditytemperature probes to a PC, either by means of a USB port or by means of a wired / wireless TCP/IP network (LAN / Internet).
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 4 of 24 HygroClip DI-4 • • • • RJ45 port (wired Ethernet LAN) Two probe inputs (HygroClip probes) RS-485 port (4-pin connector: TX +, TX - , DC in, GND) Receptacle for AC adapter • • • • Wireless TCP/IP interface Two probe inputs (HygroClip probes) RS-485 port (4-pin connector: TX +, TX - , DC in, GND) Receptacle for
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 5 of 24 HygroClip S measurement in air max. 85°C (185°F) – wire mesh filter HygroClip SP05 measurement in air ducts max. 85°C (185°F) HygroClip SC05 measurement in tight spaces max. 100°C (212°F) HygroClip HK25 / HK40 air probe measurement at high temperature HK25: max. 150°C (302°F) HK40: max.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 6 of 24 The following probes have a 7-pin connector. An adapter cable such as the T7-03-B5 is required for connection to the HygroClip DI (see Accessories). Extension cable model AC1616 is not compatible with the HygroClip DI. HygroClip IC-1 (100mm) / IC-3 (250mm) for through-wall installation max.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document Type Page Document title 6.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 8 of 24 Operating System compatibility: ● Windows XP, NT4 with SP 6a or higher ● Windows 2000 with SP 2 or higher ● Windows Server 2003 HW4 was written for the Microsoft .NET framework and requires this framework to be installed on the computer. The .
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 9 of 24 The forms that are available within the Device Manager form are listed in a tree located on the left pane of the form. To select a form, click on it with the left mouse button. Device Manager Menu Bar The Device manager menu bar is located at the top of the form.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 10 of 24 Device Information ● Device name: this text will be displayed next to the instrument model information. As far as possible use a unique device name. ● RS485 address: click on the underlined link to change the instrument address to be used in conjunction with an RS-485 network (multi-drop).
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 11 of 24 Language / Unit System ● Units: use the underlined links to the right of the form to globally change the unit system used by the instrument. If the unit system offers several choices such as In Hg or PSI for pressure, choose a unit by left clicking on the arrow to the right of each text box.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 12 of 24 Probe Inputs Clicking on the plus sign to the left of “Probe Inputs” expands the tree and displays probe inputs 1 and 2. Note: the HygroClip DI accepts HygroClip probes only.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 13 of 24 Dew point (Dp) or Frost point (Fp): depending on the selection made here, HW4 will display either the symbol Dp or the symbol Fp. The symbol Fp indicates that any value below freezing is a frost point as opposed to a dew point. When selected, the symbol Fp is also be displayed for values above freezing.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 14 of 24 IMPORTANT: All devices within an RS-485 multi-drop should use the same baud rate. Communication within the RS-485 multi-drop will not work when the devices are configured with different baud rates.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 15 of 24 To protect the device, click on the underlined link. HW4 opens the following form where a password can be entered (maximum 8 characters): Enter the password and click on the OK button. HW4 confirms that the device is now protected: © 2007; Rotronic AG IN-E-HyClip DI-V1_10.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title Instruction Manual Document Type Page 16 of 24 To unprotect the device and enable all functions, click on the underlined link. HW4 opens the password form: Enter the password and click on the OK button. HW4 confirms that the device is now unprotected.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document title 7.5 Instruction Manual Document Type Page 17 of 24 Probe adjustment Please consult the HW4 on-line instruction manual. 8.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Document Type Page Document title 9. Instruction Manual 18 of 24 Ethernet local area network ROTRONIC devices with an Ethernet (TCP/IP) interface presently use an internal module manufactured by Digi International (Digi Connect ME for a wired connection, Digi Connect WI-ME for a wireless connection).
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 19 of 24 for RS-485 slaves” is being used, HW4 automatically changes from 00 to 01 the RS-485 address of any master that has slaves attached. As a result, all masters end up with the same RS-485 network address (01).
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 20 of 24 11.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 21 of 24 12. Accessories Order Code Description HW4-E HW4 Edition software (CD ROM), single user HW4-P HW4 Professional software (CD ROM) FDA/GAMP conformance, multi user MOK-02-B5 MOK-05-B5 Adapter cable for HygroClip probes with DAT05 connector (use for inputs 2 to 7). Cable length 2 meter (6.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 22 of 24 14. Electrical Installation Guidelines Power supply wiring Heavy machinery and instrumentation should not share the same power supply wiring. If this cannot be avoided, noise filters and surge protectors should be used. Most UPS devices have those features already integrated.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 23 of 24 Lightning protection Cabling in areas with a risk of lightning requires a lightning protection. For cabling underground in between buildings, we recommend the use of special fiber optic cables. If this is not possible, use copper cables that are suitable for underground installation.
IN-E-HyClip DI-V1_10 Rotronic AG Bassersdorf, Switzerland Document code Unit HygroClip DI digital interface: instruction manual Instruction Manual Document Type Page Document title 24 of 24 The following points should also be observed: • • • • Close the parasitic circuit Connect all devices to the potential equalizing cable as often as possible. Electrical conductors such as machine elements, metal tubes or supporting constructions should be integrated into the system.