Specifications
 29 
7.6. Calibration Mode 
To CALIBRATE the scale, the span switch must be enabled. 
Indicators 
1 :   2 : NET  3 : Îgh 
INDICATORS 
DISPLAY TASK OPERATION 
1 2 3 WEIGHT MIN  MAX 
Weighing mode.   
•
••
• 
0.000 0 0 
1. Press Span Switch 
S-on   
2. To enter calibration mode, 
press & hold Re-zero while 
pressing 8, 7, 1, 5. 
[Re-zero] 
[8] [7] [1] [5] 
90614 CAL00  0.002 
3. Press Re-zero key. 
[Re-zero]   
90614 CAL00  0.000 
4. With no weight on the platter, 
press [½
½½
½]key in order to 
calibrate the zero point. 
[½
½½
½]   
- - - - -  - - - - -  - - - - - 
Span Calibration 
0.000 CALSP  87342 
5. Place capacity test weight on 
platter 
0.271 CALSP 66000 
6. Press [½
½½
½]key in order to 
calibrate the span. 
[½
½½
½]   
- - - - -  - - - - -  - - - - - 
7. After a few seconds, the scale 
shows the span counts. 
60000 SPAn CoUnt 
8. Remove the weight. Press [%] 
key to exit calibration mode. 
[%] 
•
••
• 
S-on  
9. Calibration finished. Press  
Span Switch to return to the 
weighing mode. 
•
••
• 
0.000 0 0 
7.7. Error Indication 
Overflow – Indication – weight display blanks and all range indicator lights flash. 
Underflow – Indication – weight display blanks and all range indicator lights flash. 
When scale is outside the re-zero range – Indication –Display shows all 8’s 
88888 88888 
888888 
When calibrating scale and scale is behind zero – Indication – Display shows 
PrESS 
 SPr Int 
When calibrating scale and calibration weight is incorrect – Indication – Display 
shows (incorrect value) 10000 - - - - - 
 - - - - - - 
When calibrating scale and mV/V setting is incorrect – Indication – Display shows 
(all dashes) - - - - - - - - - - 
 - - - - - - 