Specifications
Configuration Settings 29
3.2.2 Uploading Saved Spec Configurations to a DC-200/300
To download the spec settings from a DC-200/300 to the computer, see Section 3.2.1. If you already have a
saved configuration file of spec settings, use the instructions below to upload the settings to a DC-200/300. This
could be to change the settings of a DC-200/300 to fit a particular task, department or facility or it could be to
take the spec settings configuration from one DC-200/300 and upload it to other DC-200/300’s so that their
settings match.
1. With the scale turned on and at the Stand-by screen, press
REZERO + SHIFT UP. The scale display will
indicate that it is changing to the data transfer mode.
2. The scale display will next indicate that it is ready for the data transfer mode after the scale is restarted.
Turn the scale off and then on again, using the on/off switch located on the bottom of the scale at the
back.
3. After you restart the scale and it goes through its startup sequence, the scale will show that the data
transfer mode is ready.
4. On your computer, open the Spec Upload program.