Specifications
6.7.2.1. SETTING – UP SINGLE CHANNEL SCALE
IF YOU ARE SETTING UP A SINGLE CHANNEL SCALE WITH ONE LOAD CELL
THIS IS THE WAY THE SPECS 16 & 17 SHOULD BE SET.
16
∗∗
∗∗
∗∗
∗∗
∗∗
∗∗
Normally
Set
0 0 0 1
_ _ _ _
SCALE 1:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
SCALE 2:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
17
∗∗
∗∗
∗∗
∗∗
∗∗
∗∗
Normally
Set
1 0 1 1
_ _ _ _
SCALE 3:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
SCALE 4:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
The Load Cell Would Be Plugged Into The SX-1 Load Cell Connector. (See pg. 13 for wiring.)
If You Are Using The SX-L Load Cell Platform No Power Supply Will Be Necessary As Power Will Be
Supplied By The Platform.
If You Are Using Any Other Type Of Platform It Will Be Necessary To Use A PS-100 Power Supply.
In This Case Other Specs That Will Need To Be Set Are As Follows:
Weight and Measures Specification: To enter this mode, enter the numeric keys 1,4,2 while
pressing the Re-zero Key. The Span Switch must be “ON” to enter this mode.
Spec No. Bit 3 Bit 2 Bit 1 Bit 0
20
_ _ _ _
Minimum Display ( Scale 1)
00 : 2 10 : 5
01 : 1 11 : 10
!!
!!
1
Minimum Display (Scale 2)
00 : 2 10 : 5
01 : 1 11 : 10
21
_ _ _ _
Weight Decimal Point Position (Scale 1)
000 : 00000 011 : 00.000
001 : 0000.0 100 : 0.0000
010 : 000.00
!!
!!
1
23
Normally
Set
0 0 1 1
Display Resolution
00 : 1/10,000 10 : 1/2,500 !!
!!
1
01 : 1/5,000 11 : Not Used
Zero Setting Range
00 : + Unlimited 10 : +- 10% FS
- 10% FS
01 : +- 2% FS 11 : Not Avail.
25
0 0 0 0
Scale Starting Method
0 : Automatic
1 : Manual
IR Mode
protected by
Span Switch
0 : No
1 : Yes
Scale Type
0 : Single
Scale !!
!!
2
Gross Mode Available
0 : Yes
1 : No
32
1 0 1 0
Calibration Mode
protected by Span
Switch
0 : Yes 1 : No
Battery Low
Lamp
0: Yes
1: No
Auto Exit from
Add Mode
0 : No
1 : Yes
External Load Cell (Scale 3)
0: No !!
!!
2
!!
!!
1 To Select The Minimum Display And Decimal Location Divide Capacity By The Display Resolution
Selected In Spec 23.
!!
!!
2 In This Case Only Scale 1 Will Be Turned On.