Specifications
6.7.2. SETTING UP DC – 190
When setting up a DC – 190 or adding a second channel, third channel or even a
force balance, it is necessary to assign a different location for each scale whether
it is present or not.
Specs 16 & 17 only set up the location for each of the four scales. These specs do
not turn on the additional scales. Scales 2 and 3 are turned on and off in the
Weights & Measures specs (see spec 25 bit 1 & spec 32 bit 0). The Force Balance
is turned on and off in spec 8 bit 3 and the connector type is selected in spec 13
bits 2-1-0.
!!
!!
When setting up a DC-190 or adding a second or third platform. Be sure to
plug in the remote scale, simulator or dummy plug (consult factory on dummy
plug wiring), before connecting A/C to the DC-190 and before turning on Scales
2 and/or 3. Do not unplug remote scale from the DC-190 while the DC-190 is
powered.
∗∗
∗∗
∗∗
∗∗
∗∗
∗∗
SPEC 16 & 17 SHOULD ONLY BE CHANGED BY AN AUTHORIZED SERVICE TECHNICIAN !!!
16
∗∗
∗∗
∗∗
∗∗
∗∗
∗∗
Normally
Set
0 0 0 1
_ _ _ _
SCALE 1:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
SCALE 2:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
17
∗∗
∗∗
∗∗
∗∗
∗∗
∗∗
Normally
Set
1 0 1 1
_ _ _ _
SCALE 3:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
SCALE 4:
00: Internal Scale 1
01: Internal Scale 2
10: External Scale
11: Force Balance
ALL SCALES ARE UNIQUE AND EACH MUST HAVE THEIR OWN CHANNEL LOCATION.
IF YOU WISH TO USE THE REMOTE PLATFORM CONNECTOR FOR SCALE
2, IT IS NECESSARY TO CHANGE THE SCALE 3 SPEC LOCATION.
EXAMPLE : USING SINGLE SX PLATFORM SCALE: It is necessary to assign
a different location for each scale whether it is present or not.
SPEC 16 SCALE 1 = 00: INTERNAL SCALE 1 / SCALE 2 = 10 EXTERNAL SCALE
SPEC 17 SCALE 3 = 01: INTERNAL SCALE 2 / SCALE 4 = 11 FORCE BALANCE