Specifications

6.6.5.3. RS232C Related Spec List
The following R-141 spec must be set correctly in order for DC-190 to communicate with the
Force Balance
Spec No. Bit 3 Bit 2 Bit 1 Bit 0
8 RS-232C
Connection
(Force Balance)
0 : No 1 : Yes
RS-232C
Data Length
(Optional)
0 : 7 bits 1 : 8 bits
RS-232C Baud Rate
(Optional)
00 : 19200 10 : 4800
01 : 38400 11 : 9600
9 RS-232C
Stop Bit
(Optional)
0 : 1 bit 1 : 2 bits
Force Balance Type
0: SHG-300 1: TP-200
RS-232C Parity Bit
(Optional)
00 : No 10 : Not Used
01 : Odd 11 : Even
13
RS232(PC)
Header:
0: Code
1: Title
RS232 CONNECTOR
Sub Din
000: Printer Force Bal.
001: Force Bal. Printer
100: PC Force Bal.
101: Force Bal. PC
010: Printer PC (*1)
011: PC Printer (*2)
USA Default Spec:
SPEC08: 0010
SPEC09: 0111
SPEC13: 0001
Note: SPEC08 bit 3 for RS232 Connection is disable in Default Spec.
SPEC13 bit 0, 1 &2 need to be changed before using the force balance.