Specifications
DC-190 RS-232 specification (continued)
6.6.5. FORCE BALANCE
DC190 can support: OHAUS EXPLORER
6.6.5.1. General Specification
Baud Rate 1200 /2400 / 4800 / 9600 bps (Spec 8 bits 0 & 1)
Data Length 7 bits / 8 bits (Spec 8 Bit 2)
Parity None/ Odd / Even (Spec 9 Bit 0 & 1)
Stop Bit 1 bit / 2 bits (Spec 9 Bit 3)
F. Balance Type OHAUS EXPLORER (Spec 9 Bit 2)
6.6.5.2. Calibration
6.6.5.2. OHAUS EXPLORER /TP200
Preparation For EXPLORER
1. Perform Initial Setup For EXPLORER according to its Service Manual
2. Set the RS232 setting in EXPLORER
3. Set Auto Print with Continuously option.
4. Set Numeric Only to Off .
5. Set the Unit to OZ or G
Linking to DC190
1. 1.Connect two scales with cable.
2. Select Scale 4 by pressing Scale key several times.
3. Calibration is not needed
Please Note:
1. Re-zero and zero tracking are disable when using EXPLORER/TP200.
Weight Display in Dc190 will mask.
2. The negative start range for oz mode can up to around –3.00000oz of the capacity.
3. The max tare range for oz mode can up to 7oz.