Specifications
6.5. DC-190 Specification List ver.3.37 (continued)
Weight and Measures Specification : To enter this mode, enter the numeric keys 1,4,2 while
pressing the Re-zero Key. The Span Switch must be “ON” to enter this mode.
Spec No. Bit 3 Bit 2 Bit 1 Bit 0
20
_ _ _ _
Minimum Display ( Scale 1)
00 : 2 10 : 5
01 : 1 11 : 10
Minimum Display (Scale 2)
00 : 2 10 : 5
01 : 1 11 : 10
21
_ _ _ _
Eltron Printer
Selection
0 = 2722
1 =2742 or 2642
Weight Decimal Point Position (Scale 1)
000 : 00000 011 : 00.000
001 : 0000.0 100 : 0.0000
010 : 000.00
22
_ _ _ _
Weight Decimal Point Position (Scale 2)
000 : 00000 011 : 00.000
001 : 0000.0 100 : 0.0000
010 : 000.00
23
0 0 0 0
Display Resolution
00 : 1/10,000 10 : _,500
01 : 1/5,000 11 : Not Used
Zero Setting Range
00 : + Unlimited 10 : +- 10% FS
- 10% FS
01 : +- 2% FS 11 : Not Avail.
24
0 0 0 0
Masked Display
at Minus Wt.
0 : Gross
1 : Net
Display at
Minus Wt.
0 :Minus
Display
1 :Masked
Zero Lamp
Lighting
Method
0 : Gross
1 : Net
When No AC,
Display Mask
When Battery
Low or No
Battery.
0 : Yes 1 : No
25
0 0 _ 0
Scale Starting
Method
0 : Automatic
1 : Manual
IR Mode
protected by
Span Switch
0 : No
1 : Yes
Scale Type
0 : Single
Scale
1 : Double
Scale
Gross Mode
Available
0 : Yes
1 : No
26
0 0 0 0
Zero Tracking
When Tare
0 : Yes
1 : No
Weight Reset
when Tare
0 : Yes
1 : No
Initial Start Range
00 :+ Unlimited 10 : +- 10% FS
- 10% FS
01 : +- 2% FS 11 : Not Avail.
27
0 1 0 0
Comma Display
0 : No 1 : Yes
Digital Tare
Setting
0 : No 1 : Yes
Tare Range
00 : 100%FS 10 : 5% FS
01 : 50%FS 11 : Not Available
28
0 0 0 0
Auto Tare clear
when Rezero
0 : No
1 : Yes
Automatic Unit Weight Clear
Condition
00 : Over Net 5d and Gross 21d
and Weight Stable
01 : >= Net 1d and Weight Stable
10 : >= Net 1d and Quantity >0
and Weight Stable
Automatic Unit
Weight Clear
0 : No
1 : Yes
29
0 0 0 0
Di
g
ital Tare
Rounding
0 : Tare
Exactly
1 : Round to
Nearest
Increment
Tare Value
Exchange
0 : Yes
1 : No
Tare Addition
0 : Yes
1 : No
Tare
Subtraction
0 : Yes
1 : No