Specifications
18 DC-782 Operation Manual
3. The scale display will confirm that you are in the Calibration Mode.
4. Make sure that there is no weight on the platform of the sc
ale you are calibrating and press the *
CONFIRM key. The scale will flash briefly as it searches for the zero point and then will alternate
between showing CAL00 and the capacity.
5. Place a reference weight on the platform. It is preferable to
calibrate the scale using a weight equal to the
full capacity of the scale (i.e. a 15 lb weight for 15 lb capacity scale, etc.) If the reference weight is not
equal to the full capacity of the scale, it must at least be greater than 10% of the full scale capacity.
If the reference weight is equal to the full
capacity of the scale, press the * CONFIRM key. If the
reference weight is less than the full capacity of the scale, enter the value of the weight placed on the
platter using the numeric keys before pressing the * CONFIRM key. The weight you enter will appear in
the Weight display.
6. The display will flash briefly again as the span settings for cal
ibration of the scale are being temporarily
saved. Then the display will return to the weighing mode with the calibrated weight showing in the
Weight window..
7. To exit the Maintenance Mode and save your calibration, power down your DC-782 scale. To exit the
Maintenance
Mode and return to the Weighing Mode without saving your calibration, press the TARE
key.