User manual
Triggers
R&S
®
RTO
82User Manual 1316.0827.02 ─ 06
"Auto"
The holdoff time is calculated automatically based on the current hori-
zontal scale.
"Auto time scaling" defines the factor the horizontal scale is multipied
with.
"Auto time" shows the resulting holdoff time: Auto time = Auto time
scaling * Horizontal scale.
"Off"
No holdoff
SCPI command:
TRIGger<m>:HOLDoff:MODE on page 485
TRIGger<m>:HOLDoff:TIME on page 486
TRIGger<m>:HOLDoff:EVENts on page 487
TRIGger<m>:HOLDoff:MAX on page 487
TRIGger<m>:HOLDoff:MIN on page 487
TRIGger<m>:HOLDoff:AUTotime on page 488
TRIGger<m>:HOLDoff:SCALing on page 488
3.3.4.2 A - B - R
The complex trigger sequence "A → B → R" consists of two trigger events - A and B - and
an optional reset condition.
See also: chapter 3.2.4, "Setting Up a Trigger Sequence", on page 56.
Couple levels of all events
Sets the channel trigger levels of the A-, B-, and R-event to the values of the current event
(per channel).
Reference for Triggers