User manual

Triggers
R&S
®
RTO
80User Manual 1316.0827.02 ─ 06
Absolute hysteresis
Defines a range in absolute values around the trigger level. If the signal jitters inside this
range and crosses the trigger level thereby, no trigger event occurs.
SCPI command:
TRIGger<m>:LEVel<n>:NOISe:ABSolute on page 483
Relative hysteresis
Defines a range in divisions around the trigger level. If the signal jitters inside this range
and crosses the trigger level thereby, no trigger event occurs.
SCPI command:
TRIGger<m>:LEVel<n>:NOISe:RELative on page 483
Couple levels (Trigger level and hysteresis coupling)
Sets the trigger levels and hysteresis values for all channels to the values of the currently
selected trigger source. The function affects only the levels defined for the selected event
(A-, B-, or R-event). The hysteresis of the external trigger input is an independent value,
and it is not affected by level coupling.
In trigger sequences, event coupling of trigger levels is possible: "Couple levels of all
events" on page 82
SCPI command:
TRIGger<m>:SCOupling on page 456
3.3.4 Sequence
A trigger sequence consists of at least one trigger event and additional conditions defining
when the trigger occurs.
The simple sequence "A only" contains an A-event and the holdoff setting as optional
condition.
The complex trigger sequence "A B R" consists of two events - A and B - and an
optional reset condition. After the A-event conditions have been met, the B-event with
independent conditions is enabled. A- and B-events are configured in the same way.
Without any reset, the instrument waits until one or a specified number of B-events occurs
that causes the trigger, and then the sequence starts again. If you expect, for example,
an irregular B-trigger event, you can configure a reset condition to restart the sequence
with the A-event. The reset condition can be a simple timeout, or a trigger event that is
defined in the same way as the A- and B-trigger events.
The instrument checks the trigger settings for compatibility and disables settings that do
not fit the previous settings in the sequence.
See also: chapter 3.2.4, "Setting Up a Trigger Sequence", on page 56.
3.3.4.1 A Only
The "A only" sequence contains an A-event and the holdoff setting as optional condition.
T
A
T
A
B
R
Reference for Triggers