User manual

Remote Control
R&S
®
RTO
770User Manual 1316.0827.02 ─ 06
Positive error numbers are instrument-dependent. Negative error numbers are reserved
by the SCPI standard.
Return values:
<Error> List of: Error/event_number,"Error/event_description>[;Device-
dependent info]"
If the queue is empty, the response is 0,"No error"
Usage: Query only
SCPI confirmed
SYSTem:ERRor[:NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue. The
response consists of an error number and a short description of the error.
Positive error numbers are instrument-dependent. Negative error numbers are reserved
by the SCPI standard.
Return values:
<Error> Error/event_number,"Error/event_description>[;Device-depend-
ent info]"
If the queue is empty, the response is 0,"No error"
Usage: Query only
SCPI confirmed
SYSTem:ERRor:CODE:ALL?
Queries the error/event queue for all unread items and removes them from the queue.
The response is a comma separated list of error numbers in FIFO order.
Positive error numbers are instrument-dependent. Negative error numbers are reserved
by the SCPI standard.
Return values:
<Error> If the queue is empty, the response is 0
Usage: Query only
SCPI confirmed
SYSTem:ERRor:CODE[:NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue. The
response is the error number.
Positive error numbers are instrument-dependent. Negative error numbers are reserved
by the SCPI standard.
Return values:
<Error> If the queue is empty, the response is 0
Usage: Query only
SCPI confirmed
Command Reference