User manual
Remote Control
R&S
®
RTO
761User Manual 1316.0827.02 ─ 06
EXPort:WAVeform:STARt <Start>
Sets the start value of the waveform section for export, if EXPort:WAVeform:SCOPe is
set to Manual.
Parameters:
<Start> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Default unit: s
Firmware/Software: V 1.25
EXPort:
WAVeform:STOP <Stop>
Sets the end value of the waveform section for export, if EXPort:WAVeform:SCOPe is
set to Manual.
Parameters:
<Stop> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Default unit: s
Firmware/Software: V 1.25
EXPort:WAVeform:ZOOM <DiagramName>, <ZoomName>
EXPort:WAVeform:ZOOM? <DiagramName>
Sets the zoom area to be used for limited data export if EXPort:WAVeform:SCOPe is
set to ZOOM.
Parameters:
<ZoomName> Name of the zoom diagram
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
Firmware/Software: V 1.25
EXPort:WAVeform:CURSorset <Cursorset>
Sets the cursor set to be used for limited data export if EXPort:WAVeform:SCOPe is
set to CURSor.
Parameters:
<Cursorset> CURSor1 | CURSor2 | CURSor3 | CURSor4
*RST: CURSor1
Firmware/Software: V 1.25
Command Reference