User manual

Remote Control
R&S
®
RTO
759User Manual 1316.0827.02 ─ 06
Example:
MMEM:ATTR? 'C:\USER\DATA\*.*'
"Datei1.LOG,A", "Datei2.LOG,A",
"Datei3.LOG,ASH", "Datei4.DLL,RSH",
"Datei5.INI,SH"
MMEMory:STORe:STATe <MemoryNumber>, <FileName>
Stores the instrument settings from the specified internal memory to the specified file. To
store the current instrument settings to the internal memory, use *SAV first.
Setting parameters:
<MemoryNumber> Number of the internal memory
Range: 1 to 99
<FileName> String parameter specifying the complete path and filename of the
source file.
Example:
*SAV 4
MMEM:STORe:STATe 4,'C:
\Settings\Settings_1051.dfl'
Saves current instrument settings to the internal memory number
4. Then stores the settings from the internal memory number 4 to
the file C:\Settings\Settings_1051.dfl.
Usage: Setting only
MMEMory:LOAD:STATe <MemoryNumber>, <FileName>
Loads the instrument settings from the specified file to the specified internal memory.
After the file has been loaded, the settings must be activated using a *RCL command.
Setting parameters:
<MemoryNumber> Number of the internal memory
Range: 1 to 99
<FileName> String parameter specifying the complete path and filename of the
source file.
Example:
MMEM:LOAD:STATe 4,'C:
\Settings\Settings_1051.dfl'
*RCL 4
Loads instrument settings from the file C:\Settings\Set
tings_1051.dfl to the internal memory number 4, and then
activates the settings in internal memory number 4.
Usage: Setting only
16.2.16.2 Data Export
EXPort:WAVeform:SOURce............................................................................................760
EXPort:WAVeform:SCOPe..............................................................................................760
EXPort:WAVeform:STARt...............................................................................................761
Command Reference